Infant Incubator
Infant Incubator is one of the important equipment used in medical institutions. It is mainly used for constant temperature cultivation, temperature resuscitation, oxygen delivery, rescue, hospitalization and observation of sick infants and premature infants. It is used in hospitals, confinement centers, neonatology, etc.
Features:
- LCD display
- Double control modes of bed temperature and skin temperature
- Double-layer thermostatic cover, automatic air curtain device
- Humidity control function
- Air & Oxygen blender
- Oxygen supply system
- Baby cot is infinitely adjustable
- 24 kinds of sound and light alarm, comprehensive protection of baby safety
General
Parameters |
Item
Power supply |
Specifications
1320VA; 11 OV 60Hz (Optional) |
Size
Maximum heater output power (220VAC) |
mm
Air chamber heater: 330W, Water tank heater: 155W |
|
Auxiliary network power output
Heater indication |
AC220V/50Hz,Both maximum permissible currents are 220V I .5A
10% increasing amount |
|
Temperature control mode | Bed temperature mode;The skin temperature mode | |
Bed temperature control temperature range | 25.W37.ooc;
37.1 -039.ooc (Span Mode) |
|
Skin temperature control range | 34.W37.ooc; 37.W38.ooc (Span Mode) | |
Skin temperature sensor temperature display range | 5.w65.ooc | |
Heating time | s 50min | |
Temperature variability | ||
Temperature uniformity
(mattress in horizontal position) |
0.80C |
Humidity Control | The running time of the tank after it is filled with distilled water | Humidity below 70%RH, at least 24h |
Water tank capacity
Humidity display range |
1500ml | |
Humidity control range
Humidity control accuracy |
% per adjustment) | |
Oxygen
Concentration Control |
Oxygen concentration display range | 00W99% |
Oxygen concentration display resolution | 1% | |
Oxygen concentration control accuracy | ±5% oxygen volume concentration | |
Oxygen concentration setting range | 21% — 60% (1% per adjustment) | |
Oxygen sensor service life | The maximum use time is 10000h at 100% oxygen concentration | |
Transport Storage | Temperature | -200C&+550C |
Atmospheric pressure range | 500hPaæ.1060hPa | |
Relative humidity | 93%RH | |
Packing size/ Gross Weight
Relative humidity |
Base packing size: 50Kg
Body packing size: 1100*710*855mm 60Kg |
|
The Work
Environment |
The environment temperature | |
Pressure range | 700hParv1060hPa | |
Service altitude | 9000m | |
Over voltage category | Il | |
Pollution levels | 2 | |
Ambient air velocity | s 0.3m/s | |
Other
Indicators |
Environmental noise in incubator baby compartment | No more than 55dB(A) at stable temperature |
Carbon dioxide concentration in the infant compartment | A mixture of 4% carbon dioxide and air was fed into an 8mm diameter tube in the vertical direction from the mat to the top of the box at 750mL/min to a point 1 Ocm above the mattress center and IOcm above the center. The concentration was less than 0.5% after stabilization | |
The velocity of air flow over a baby’s mattress | ‘0.35m/s | |
General
Alarm |
Power off alarm,Fan alarm, sensor alarm, Deviation alarm, Over temperature alarm, Water tank placement error alarm, Water shortage alarm | |
The System
Alarm |
Mcul and Mcu2 communication error alarm, Mcu2 and Mcu3 communication error alarm, Analog-to-digital conversion chip 2544 fault alarm, Analog-to-digital conversion chip 2548 fault alarm, Over temperature alarm, Temperature heating system fault, Humidity heating system fault |
Model
Temp. Control |
Type l,
Bed/Skin temperature |
Type I l
Bed/Skin temperature |
Display | LCD | LCD |
Bed Temp. Control Range | 250C-390C | 250C-390C |
Skin Temp. Control Range Humidity Control Range | 340C-380C | |
Oxygen Control Range | ||
Cabinet | Yes | Yes |
Compare
Reviews
0.0 Average Rating Rated ( 0 Review )
There are no reviews yet.